CLIA Kit for Growth Differentiation Factor 3 (GDF3)
MatrixRecovery range (%)Average(%)
serum(n=5)82-10486
EDTA plasma(n=5)80-9279
heparin plasma(n=5)80-9285
Request to COA Click Here

Description

Specificity

This assay has high sensitivity and excellent specificity for detection of Growth Differentiation Factor 3 (GDF3).
No significant cross-reactivity or interference between Growth Differentiation Factor 3 (GDF3) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Growth Differentiation Factor 3 (GDF3) and the recovery rates were calculated by comparing the measured value to the expected amount of Growth Differentiation Factor 3 (GDF3) in samples.

MatrixRecovery range (%)Average(%)
serum(n=5)82-10486
EDTA plasma(n=5)80-9279
heparin plasma(n=5)80-9285

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Growth Differentiation Factor 3 (GDF3) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Growth Differentiation Factor 3 (GDF3) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Growth Differentiation Factor 3 (GDF3) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample1:21:41:81:16
serum(n=5)91-99%84-96%85-97%87-101%
EDTA plasma(n=5)87-103%97-104%78-97%80-93%
heparin plasma(n=5)78-99%93-101%91-101%81-92%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

ReagentsQuantityReagentsQuantity
Pre-coated, ready to use 96-well strip plate1Plate sealer for 96 wells4
Standard2Standard Diluent1×20mL
Detection Reagent A1×120µLAssay Diluent A1×12mL
Detection Reagent B1×120µLAssay Diluent B1×12mL
Substrate A1×10mLSubstrate B1×2mL
Wash Buffer (30 × concentrate)1×20mLInstruction manual1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Additional information

Catalog No.

SCC109Mu

Pack Size

48T, 96T, 96T*10, 96T*100, 96T*5